Modification, reconstruction and commissioning of a vapour recirculation apparatus for high-pressure low-temperature vapour-liquid equilibrium measurements.

UKZN ResearchSpace

Show simple item record

dc.contributor.advisor Ramjugernath, Deresh D.
dc.creator Knock, Jason.
dc.date.accessioned 2011-10-12T06:38:21Z
dc.date.available 2011-10-12T06:38:21Z
dc.date.created 2010
dc.date.issued 2010
dc.identifier.uri http://hdl.handle.net/10413/3762
dc.description Thesis (M.Sc.Eng.)-University of KwaZulu-Natal, Durban, 2010. en
dc.description.abstract The vapour recirculation apparatus of Moodley [2002] was modified, reconstructed, and commissioned in order to be used in the acquisition of high-pressure vapour-liquid equilibrium (HPVLE) data at low temperatures. The original equipment of Moodley [2002] was modified with the aim of achieving the desired operating temperature range which the original equipment was unable to achieve. Major modifications were carried out on the cooling circuit allowing the equipment to reach temperatures as low as -30°C, a significant improvement to the original equipments minimum attainable temperature of -5°C. Modifications were also successfully carried out on the vapour recirculation pump of Moodley [2002], which failed when operated over extended periods at high pressures thus enabling the equipment to operate at pressures up to 10.0 bar, an improvement on the previous pressure operating limit of 6.9 bar. The operating limits of the equipment were tested through measurement of pure-component vapour-pressures of isopentane at temperatures between -14 and +27.9°C and pressures up to 10.1 bar and on propane at temperatures between -30.1 and +26.0°C and pressures up to 9.7 bar. The isopentane vapour-pressure measurements had an average deviation of ±0.49% when compared to literature data while the propane vapour-measurements had a maximum average deviation of ±0.35% when compared to literature data indicating that the equipment was capable of measuring accurate vapour-pressure data at temperatures down to -30°C and pressures up to 10.0 bar. The equipment was thereafter used in the acquisition of binary HPVLE data. Considerable time was spent developing and practicing the techniques used in the binary HPVLE measurements. Binary measurements were performed on the test system propane + 1- propanol at 19.9°C. To gain more confidence in the binary HPVLE measurements another test system, propane + isopentane was selected and binary HPVLE measurements were performed at 25°C and 0°C. The equipment was able to reproduce relatively accurate binary HPVLE results for the test systems at the selected isotherms. The equipment was thereafter used in the acquisition of a new set of binary HPVLE data for the propane + isopentane system at -10°C however owing to time constraints and chemical availability the acquisition of a complete set of data was not possible. The binary HPVLE data was thereafter regressed via the direct method. The Peng-Robinson (PR) equation of state (EOS) and the Soave-Redlich-Kwong (SRK) EOS were each coupled with the Mathias Copeman alpha function together with the Wong-Sandler mixing rule and the NRTL local composition model and applied to the binary systems at each of the isotherms investigated. Regressed data showed a relatively good agreement with measured experimental data for both binary systems investigated at all of the isotherms except the new -10°C isotherm of the propane + isopentane system. en
dc.language.iso en en
dc.subject Vapour-liquid equilibrium--Measurement. en
dc.subject Theses--Chemical engineering. en
dc.title Modification, reconstruction and commissioning of a vapour recirculation apparatus for high-pressure low-temperature vapour-liquid equilibrium measurements. en